Evaluating stereotactic accuracy with patient-specific MRI distortion corrections for frame-based radiosurgery

Evidence

J Appl Clin Med Phys. 2024 Jul 23:e14472. doi: 10.1002/acm2.14472. Online ahead of print.

ABSTRACT

PURPOSE: This study examines how MRI distortions affect frame-based SRS treatments and assesses the need for clinical distortion corrections.

METHODS: The study included 18 patients with 80 total brain targets treated using frame-based radiosurgery. Distortion within patients’ MRIs were corrected using Cranial Distortion Correction (CDC) software, which utilizes the patient’s CT to alter planning MRIs to reduce inherent intra-cranial distortion. Distortion was evaluated by comparing the original planning target volumes (PTVORIG) to targets contoured on corrected MRIs (PTVCORR). To provide an internal control, targets were also re-contoured on uncorrected (PTVRECON) MRIs. Additional analysis was done to assess if 1 mm expansions to PTVORIG targets would compensate for patient-specific distortions. Changes in target volumes, DICE and JACCARD similarity coefficients, minimum PTV dose (Dmin), dose to 95% of the PTV (D95%), and normal tissue receiving 12 Gy (V12Gy), 10 Gy (V10Gy), and 5 Gy (V5Gy) were calculated and evaluated. Student’s t-tests were used to determine if changes in PTVCORR were significantly different than intra-contouring variability quantified by PTVRECON.

RESULTS: PTVRECON and PTVCORR relative changes in volume were 6.19% ± 10.95% and 1.48% ± 32.92%. PTVRECON and PTVCORR similarity coefficients were 0.90 ± 0.08 and 0.73 ± 0.16 for DICE and 0.82 ± 0.12 and 0.60 ± 0.18 for JACCARD. PTVRECON and PTVCORR changes in Dmin were -0.88% ± 8.77% and -12.9 ± 17.3%. PTVRECON and PTVCORR changes in D95% were -0.34% ± 5.89 and -8.68% ± 13.21%. The 1 mm expanded PTVORIG targets did not entirely cover 14 of the 80 PTVCORR targets. Normal tissue changes (V12Gy, V10Gy, V5Gy) calculated with PTVRECON were (-0.09% ± 7.39%, -0.38% ± 5.67%, -0.08% ± 2.04%) and PTVCORR were (-2.14% ± 7.34%, -1.42% ± 5.45%, -0.61% ± 1.93%). Except for V10Gy, all PTVCORR changes were significantly different (p < 0.05) than PTVRECON.

CONCLUSION: MRIs used for SRS target delineation exhibit notable geometric distortions that may compromise optimal dosimetric accuracy. A uniform 1 mm expansion may result in geometric misses; however, the CDC algorithm provides a feasible solution for rectifying distortions, thereby enhancing treatment precision.

PMID:39042450 | DOI:10.1002/acm2.14472

Document this CPD Copy URL Button

Google

Google Keep

LinkedIn Share Share on Linkedin

Estimated reading time: 6 minute(s)

Latest: Psychiatryai.com #RAISR4D Evidence

Cool Evidence: Engaging Young People and Students in Real-World Evidence

Real-Time Evidence Search [Psychiatry]

AI Research

Evaluating stereotactic accuracy with patient-specific MRI distortion corrections for frame-based radiosurgery

Copy WordPress Title

🌐 90 Days

Evidence Blueprint

Evaluating stereotactic accuracy with patient-specific MRI distortion corrections for frame-based radiosurgery

QR Code

☊ AI-Driven Related Evidence Nodes

(recent articles with at least 5 words in title)

More Evidence

Evaluating stereotactic accuracy with patient-specific MRI distortion corrections for frame-based radiosurgery

🌐 365 Days

Floating Tab
close chatgpt icon
ChatGPT

Enter your request.

Psychiatry AI RAISR 4D System Psychiatry + Mental Health